Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
Ссылки
- Third Party AdvisoryVDB Entry
- Third Party AdvisoryUS Government Resource
- Third Party AdvisoryVDB Entry
- Third Party AdvisoryUS Government Resource
Уязвимые конфигурации
Одновременно
Одновременно
Одновременно
Одновременно
EPSS
6.5 Medium
CVSS3
3.3 Low
CVSS2
Дефекты
Связанные уязвимости
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.