Логотип exploitDog
Консоль
Логотип exploitDog

exploitDog

nvd логотип

CVE-2020-10627

Опубликовано: 01 дек. 2021
Источник: nvd
CVSS3: 7.3
CVSS3: 8.1
CVSS2: 4.8
EPSS Низкий

Описание

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Уязвимые конфигурации

Конфигурация 1

Одновременно

cpe:2.3:o:insulet:omnipod_insulin_management_system_firmware:-:*:*:*:*:*:*:*

Одно из

cpe:2.3:h:insulet:omnipod_insulin_management_system:19191:*:*:*:*:*:*:*
cpe:2.3:h:insulet:omnipod_insulin_management_system:40160:*:*:*:*:*:*:*

EPSS

Процентиль: 25%
0.00085
Низкий

7.3 High

CVSS3

8.1 High

CVSS3

4.8 Medium

CVSS2

Дефекты

CWE-284
NVD-CWE-Other

Связанные уязвимости

CVSS3: 8.1
github
около 4 лет назад

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

EPSS

Процентиль: 25%
0.00085
Низкий

7.3 High

CVSS3

8.1 High

CVSS3

4.8 Medium

CVSS2

Дефекты

CWE-284
NVD-CWE-Other