Логотип exploitDog
Консоль
Логотип exploitDog

exploitDog

nvd логотип

CVE-2020-25183

Опубликовано: 14 дек. 2020
Источник: nvd
CVSS3: 8
CVSS3: 8.8
CVSS2: 5.8
EPSS Низкий

Описание

Medtronic MyCareLink Smart 25000 contains

an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

Уязвимые конфигурации

Конфигурация 1

Одновременно

cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:medtronic:mycarelink_smart_model_25000:-:*:*:*:*:*:*:*

EPSS

Процентиль: 29%
0.00103
Низкий

8 High

CVSS3

8.8 High

CVSS3

5.8 Medium

CVSS2

Дефекты

CWE-287
CWE-287

Связанные уязвимости

CVSS3: 8.8
github
больше 3 лет назад

Medtronic MyCareLink Smart 25000 all versions contain an authentication protocol vuln where the method used to auth between MCL Smart Patient Reader and MyCareLink Smart mobile app is vulnerable to bypass. This vuln allows attacker to use other mobile device or malicious app on smartphone to auth to the patient’s Smart Reader, fools the device into thinking its communicating with the actual smart phone application when executed in range of Bluetooth.

EPSS

Процентиль: 29%
0.00103
Низкий

8 High

CVSS3

8.8 High

CVSS3

5.8 Medium

CVSS2

Дефекты

CWE-287
CWE-287