Логотип exploitDog
Консоль
Логотип exploitDog

exploitDog

nvd логотип

CVE-2022-41627

Опубликовано: 27 окт. 2022
Источник: nvd
CVSS3: 4.8
CVSS3: 7.6
EPSS Низкий

Описание

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

Ссылки

Уязвимые конфигурации

Конфигурация 1

Одновременно

cpe:2.3:o:alivecor:kardiamobile_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:alivecor:kardiamobile:-:*:*:*:*:*:*:*
Конфигурация 2

Одновременно

cpe:2.3:o:alivecor:kardiamobile_6l_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:alivecor:kardiamobile_6l:-:*:*:*:*:*:*:*
Конфигурация 3

Одновременно

cpe:2.3:o:alivecor:kardiamobile_card_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:alivecor:kardiamobile_card:-:*:*:*:*:*:*:*

EPSS

Процентиль: 2%
0.00015
Низкий

4.8 Medium

CVSS3

7.6 High

CVSS3

Дефекты

CWE-311
CWE-319

Связанные уязвимости

CVSS3: 7.6
github
больше 2 лет назад

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

EPSS

Процентиль: 2%
0.00015
Низкий

4.8 Medium

CVSS3

7.6 High

CVSS3

Дефекты

CWE-311
CWE-319