Описание
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Ссылки
EPSS
7.5 High
CVSS3
Дефекты
Связанные уязвимости
The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
Уязвимость микропрограммного обеспечения медицинских устройств мониторинга состояния пациента CMS8000 Patient Monitor и Epsimed MN-120, связанная с отправкой запросов на жестко закодированный внешний IP-адрес, позволяющая нарушителю обойти ограничения безопасности и загружать и перезаписывать файлы на устройстве
EPSS
7.5 High
CVSS3